4000-cms

The Curlin 4000CMS (Central Management System), is a technologically advanced, multi-therapy ambulatory electronic infusion pump.

Description

The Moog Curlin 4000 CMS infusion pump is an easy­-to-­use, ambulatory or pole-mounted infusion device designed to meet the needs of the hospital or alternate site patient. It operates accurately in any position, has a low rate resolution of 2 microliters, and its small, compact, lightweight design allows mobility for ambulatory patients. The Curlin Medical pump can be carried in any of the four convenient carryall soft­packs, pole mounted, placed into the locked “safety shells,” or set on a suitable surface for use. Features: Cost effective, safe and easy­-to-­load disposable Curlin Medical administration sets featuring a unique, spring activated, self-­clamping, set-­based “Integral Flow-Stop” device that automatically clamps the tubing when the door of the pump is opened and prevents inadvertent fluid flow. User­-friendly, easy­-to­-teach programming that shortens staff in service time and patient teaching time. Three tamper-­resistant lock settings to assist in maintaining patient compliance and safety. Selection by user from three units of delivery, milliliters (ml), milligrams (mg), and micrograms (mcg). Retention of programmed infusion settings until cleared by the clinician. Mandatory requirement of two independent key press actions before a running pump can be turned off, thus minimizing the accidental interruption of a therapy in progress. Audio alarms that can be adjusted from 1 (very quiet) to 9 (loudest) to meet specific patient and clinical setting needs, but that cannot be totally disabled. Retained memory features that allow the pump to easily resume a therapy exactly from where it left off when it is interrupted before completion. Pump will not permit the user to enter any programming value that is outside of its predetermined range. Integral High Upstream pressure alarm to notify user of excessive upstream pressure.
Specifications
Pumping Mechanism:   Curvilinear peristaltic finger action 
Dimensions:   5 H x 4 W x 2.5 D in. 
Weight:   17.5 oz 
Case Material:  UL rated flame­-retardant, non­-toxic high impact plastic 
Operating Temperature: 
  • ± 6% Accuracy: 55° to 90°F
  • ±10% Accuracy: 45° to 55°F
  • ±10% Accuracy: 90° to 104°F
Operating Relative Humidity:  15 to 95% non­condensing
Storage Temperature Range: ­ -20° to 140°F
Storage Relative Humidity:  5 to 95% non­-condensing
Power Requirements:  2 “C size” Single­-Use Batteries or AC Power
Operating Time with Full Charge Batteries 
  • Nominal 10 hours at a rate of 400 ml/hr
  • Nominal 30 hours at a rate of 125 ml/hr
  • Nominal 85 hours at a rate of 2 ml/hr
Internal Memory Battery: 5 years life
Display:  Graphic LCD, 4 text fields, 16 characters per field
Flow Rates:  0.1 to 400 ml/hr

KVO Rates: 

0 to 10 ml/hr. Can be set at “0” in Continuous, TPN, Variable and Intermittent Therapies

Volume Accuracy:

± 6% for volume delivery >1 ml under the following conditions:

  • Temperature 55° to 90° F
  • Back pressure to 300 mmHg (5.8 PSI)

±10% for volume delivery >1 ml under the following conditions:

  • Temperature 45° to 55°F
  • Temperature 90° to 104°F
  • Back pressure to 750 mmHg (14.5 PSI)

Recommended administration set usage not to exceed 72* hours.

Administration Sets:

Proprietary Curlin Medical non­DEHP, PVC administration sets available in different configurations.

Fill Volume for Administration Sets: 

Nonfiltered microbore sets ± 4 ml. See individual administration set label for specific fill volume

 Up Occlusion Detection Time:

Microbore tubing with occlusion below spike

  • < 4 minutes at 1 ml/hr
Down Occlusion Detection Time: Microbore tubing with occlusion at distal Luer
 
Rate Low Pressure High Pressure
0.1 ml/hr  
1 ml/hr  
10 ml/hr  
50 ml/hr < 5 seconds  
125 ml/hr < 3 seconds < 5 seconds
 Programmable Time (HH:MM):
Continous 00:01 to 288:00 HH:MM
TPN 00:01 to 288:00 HH:MM
Intermittent 00:01 to 288:00 HH:MM
Variable 00:01 to 288:00 HH:MM
PCA No time limit
 Bolus Volumes:

Microbore tubing with occlusion at distal Luer

  • Low Pressure:
  • High Pressure: