Sigma Spectrum

Applications

  • Large and smal volume parenterals

  • Neonatal


Administration Set
Compatible with standard gravity IV sets supplied by Baxter and Hospira and several other companies.

Description

The Spectrum and Spectrum with Master Drug Library is intended to be used for the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products, and mixtures of required patient therapy. The intended routes of administration consist of the following clinically acceptable routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluid space. The Spectrum is intended to be used in conjunction with legally marketed intravenous administration sets and medications provided by the user.

The Spectrum and Spectrum with Master Drug Library are suitable for many user facility applications such as, but not limited to, hospitals, outpatient care areas, home care, and ambulatory care services.

The Spectrum and Spectrum with Master Drug Library are intended to reduce operator interaction through guided programming, thereby helping to reduce errors associated with complex device programming. Parameter programming requires trained health care professional confirmation of limits and drug therapy to physician’s directive.

Specifications

Dimensions:  

6.3” H x 6.4” W x 4.7” D with pole clamp and wireless battery 
Weight:   2 lbs and 2.5 oz. 

Display:

 Color, 240 x 270, LED Front-­Lit, 0.2235 mm X 0.2235 mm dot pitch

 
Pumping Mechanism:   Linear peristaltic 
AC Power:

Input 100VAC - 240VAC, 50-­60 Hz/200 mA
Output (P/N 35727): 9VDC/1200 mA, short circuit protected
Output (P/N 35714): 9VDC/800 mA, short circuit protected

Battery Power & Capacity ­ Wireless Battery Module (802.11b/g):

 

Lithium Ion, 1800 mA/h, 7.4 VDC nominal (Part No: 35162)
Capacity 4 hrs (at 125 mL/hr at the highest backlight settings)
16 hr. recharge time. Charging occurs if AC Power Adaptor is plugged in,    whether pump is ON or OFF 

Alarm Volume: 3 levels – high, medium, and low
Alarms & Alerts:  Air­-in­-line, Downstream Occlusion, Very Low Battery, Due for inspection
Logging Memory:  10,000 ± event capacity; Separate pump history log and drug event log
Anti­-Free-­Flow System: Set based, utilizing IV set slide clamp.
Dose Modes: Over 20 Modes, including: mL/hr, mg/hr, mcg/hr, mg/kg/hr, mcg/min

Flow Rate:   

0.5 to 999 mL/hr with 0.1 mL/hr increments from 0.5 to 99.9 mL/hr and 1.0 mL/hr increments from 100 to 999 mL/hr

Infusion Modes:  Primary and Secondary, Multi­-Step, and Cyclic TPN

KVO: 

The default KVO rate is set at 1 mL/hr but may be configured to between 0.5 – 50 mL/hr. 

Maximum Pump Pressure:  28 PSI
Occlusion Pressure: High (19 ±9 PSI), Medium (13 ±6 PSI), and Low (6±4 PSI)

Operational Conditions: 

Operating temperature: 60 to 80°F (15.6 to 26.7° C), 20 to 90% relative humidity, non-­condensing

Storage Temperature:

 Storage temperature: -­4 to 120°F (­-20 to 49°C), 10 to 90% relative humidity, non­-condensing

Total Drug Delivered:

 

 0.1 to 9999 mL with 0.1 mL increments from 0.1 to 99.9 mL and 1.0 mL increments from 100 to 9999 mL

 
Volumetric Accuracy (using Baxter Standard IV Sets): ±0.1mL/hr at 0.5 – 1.9 mL/hr
±5% at 2.0 – 999 mL/hr 
Wireless Network Interface (w/Wireless Battery Module (802.11b/g):  Standard: IEEE 802.11b/g
Transmit Power: 12 dBm typical
 

 

 
Features
  • Intercompatability with the Electronic Medical Record (ERM) to help improve clinical workflow and data documentation accuracy 
  • A Dose Change alert can catch potential programming errors
  • Average 96% drug library compliance that maximizes use of the key safety feature
  • Recognized programming ease-­of-­use that is designed to guide users and help minimize the number of programming errors
  • Large font with color display; easy­to­read screen with on­screen guidance
  • Standard, 802.11b Wireless, and 802.11 b/g Wireless Battery Modules available
  • The built-­in asset tracking capabilities will help quickly locate pumps for use and manage  pump inventory when used in conjunction with an asset management solution