The Ross Flexiflo Patrol incorporates the use of a specially-­designed administration set which automatically clamps off the flow of fluid when removed from the pump, preventing free­-flow. Features include a dose limit, accumulated volume counter, programmable hold time, lock-­out, and memory which retains the rate, dose, and accumulated volume. The Patrol is recommended for pediatric patients.

Specifications
Power:  
Voltage: 100 to 130 VAC, 60 Hz, 1-Phase 
Power: 15 watts, average
 Fuse: 200 mA, time-delay type
Line Cord Hospital-grade, nondetachable (10 feet)
Leakage: Less than 100 microamps
   
Mechanical:  
Height: 8.5 inches
Width: 6.5 inches
Depth: 4.8 inches
Weight: 6.6 lbs
   
Operational Specifications — Flow Rates:
Range: 1-300 mL/hr
Increments: 1 mL/hr
Accuracy: ±10% or ±0.5 mL/hr, whichever is greater
Pressure: Pumps against 18 psi nominal back pressure before no FLO alarm. Refer to PATROL® MAINTENANCE AND REPAIR
MANUAL for test procedure.
   
Battery Operation:
Type: Rechargeable sealed lead acid
Voltage: 6 V; the pump will fully recharge in a maximum of 4.5 hours while pump is off, or 12 hours if the pump is operating.
Expected Life: When fully charged, 8 hours at 125 mL/hr.
 

Standards:
Designed and manufactured to meet requirements of CSA Standard C22.2 No.125