Outlook 100es

The Outlook® ES is intended for use with B Braun Medical Inc. Horizon® Pump IV Sets to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipids, and Total Parenteral Nutrition (TPN).

Description

The Outlook® ES is intended for use with B Braun Medical Inc. Horizon® Pump IV Sets to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipids, and Total Parenteral Nutrition (TPN).

Positive pressure is frequently a necessity, but clinical experience shows that high pressure limits may increase the severity of an infiltration without causing an alarm. Because there is a need to control pump pressure settings, the Outlook ES has user-selectable Occlusion Limit settings which start at 75 mmHg and extend to 500 mmHg. For epidural infusions which require higher pressures, the pump has an extended occlusion limit setting of 750 mmHg.

The pump is equipped with distinct audible and visual alarm signals to indicate Keep Vein Open (KVO), low battery, and other alarm conditions.

Specifications
Height:  4.5 in (11.4 cm) 
Width:   12 in (30.5 cm) 
Depth:   8.75 in (22.2 cm) 
Weight:   11 lbs (5 kg) 
Type of Unit: Volumetric (Non Drop­-Counting), Positive Pressure Displacement Reservoir
Classification: Class 1-­Internally Powered, CF type, IPX1 drip-­proof, not suitable in the presence of flammable anesthetic and intended for continuous use in accordance with UL2601­1.
Moisture Protection:  IPX1; drip protected for horizontal usage
EMI/EMC Compliance: Meets or exceeds EN60601-­1-­2:2001 for Class B, emissions and immunities standard.
Ground Impedance: Less than 0.2 ohms (tested per UL2601­-1)
Power:  120 VAC, 2A, 60 Hz excluding accessories
Grounding Resistance:  Meets UL Standard 2601­-1
Leakage current: Less than 300 microamps earth leakage (tested per UL 2601­-1)
Plug:  Hospital Grade (3 pin)
Pole Size Range:  0.875" - 1.25"
Battery: 12 V sealed lead­acid battery
Battery Operating Time:  7 hours at 25 mL/hr
Mechanism:  Volumetric, Positive Pressure Displacement Reservoir
Rate Range:  0.1 to 999.9 mL/hr in 0.1 mL increments
Micro Mode: 0.1 to 99.9 mL/hr in 0.1 mL increments
Volume to be Infused Range:  0.1 to 9999.9 mL in 0.1 mL increments
Variable Occlusion Pressure:  75, 100, 200, 300, 400, 500, and 750 mmHg
KVO Rate:   3 mL/hr or selected rate whichever is lower
Volumetric Accuracy:  ± 5%
Alarms: Air-­in-­Line, Bag Empty, Door Open, Downstream Occlusion, Hold time Exceeded, Low Battery, Plug-­in, System Error, Upstream Occlusion, Check Set, DoseGuardTM Alert
Memory:  Permanent data retention after the pump is turned off unless alternative
Tubing: Horizon® Pump Sets
Modes/Features:  Dose Mode with Drug Menu and DoseGuardTM dose limits, Primary and Piggyback Modes, Set Time and Volume, Schedule Next Run, Profile Mode for TPN, Program Mode, q(x) Mode, Site Trend, Variable Occlusion Pressure, Micro Mode, Epidural Mode
Features
  • Promotes patient safety by helping to avoid accidental under-­ or overdosing using audible and visual alerts for soft/hard dose limits.
  • Provides the ability to customize up to 300 drugs and 15 care areas to match your drug formulary, using the expanded DoseGuardTM drug library.
  • Provides reliable operation and accuracy using a proven delivery mechanism.
  • Provides comprehensive “at-­a­-glance” viewing of infusion status on a large color screen.
  • Promotes quick and easy drug selection using an active alphabetic drug library matrix.
  • Promotes safer and more confident programming with color­-coded key pad and audible key clicks.