Kangaroo Joey

The Kangaroo Joey™ enteral feeding pump provides feeding and flushing in a completely portable device. This feeding pump is the first hydrating and ambulatory enteral device available. It assures clinicians of accurate and reliable function while providing patients with freedom and mobility.

Description

The Kangaroo Joey Enteral Feed and Flush Pump with Pole Clamp is a simple­to­use, precision enteral feeding pump. It is easily programmed to provide patients with eithercontinuous or intermittent feeding, and can provide automatic flushing capability when used with KangarooTM Joey Feed & Flush Sets.

Specifications
Type Infusion Device:  Volumetric 
Pumping Mechanism: Rotary Peristaltic
Pump Sets: KangarooTMJoey Feed­-Only Set or Feed & Flush Set
Feeding Formula Delivery Rate: 1-­400 mL/hr in 1 mL increments
Feeding Formula VTBD: 1­-3000 mL in 1 mL increments
Bolus Volume: 1­-3000 mL in 1 mL increments
Number of Boluses:  1­-99
Bolus Interval: 1-­24 hours in 1­hour increments
Flushing Solution Dose Range: 10-­500 mL in 1 mL increments
Flushing Solution Interval Range: 1-­24 hr in 1 hr increments
Accuracy:

7% or 0.5 mL/hr, whichever is larger, with top of the fluid column at 15.24 cm (6”) ± 0.76 cm (0.3”) above the top of the closed pump door, at a room temperature of 22°C ± 2°C (72°F ± 3°F), using a new Kangaroo Joey feeding set for no longer than the recommended 24 hours of maximum usage.

Occlusion Pressure: 15 psi (103 kPa) Nominal
Dimensions:

Height: 10.4 cm (4.1”) Width: 13 cm (5.1”) Depth: 9.1 cm (3.6”) Weight: 0.77 kg (1.7 lbs), 1 kg (2.3 lbs) with pole clamp

Material: 

•Soft-­Touch Coating: Latex­-Free Thermoplastic Urethane
•Housing: Polyester/ Polycarbonate blend

Power: Use AC power adapter for wall outlet usage. Pump operates on 5V DC, 2.4 Amps.
Battery:

New, fully charged NiMH Battery Pack delivers ≥ 18 hrs at 125 mL/hr feed rate. Approximately 15 minutes prior to complete battery discharge, a low battery alarm will occur.

Alarms: 

• System Error
• Flow Error
• Pump Set Dislodged Error
• Hold Error
• Rotor Error
• Feed Error
• Flush Error
• Battery Low

Alarm Volume: Minimum of 65 dBA at 1 meter in maximum volume orientation
Operating Temperature: 10°­ - 40° C (50°-­104° F) 75% R.H. non-­condensing
Transport & Storage Temperature: 0° - 50° C (32° - 122° F) 95% R.H. non­-condensing
   
Features
  • MISTIC (Magnetic Intelligent Set Type Identification Connector) Feed­-Only vs. Feed & Flush Set identification system to ensure match between the pump’s user interface and Pump Set type
  • Automatic Anti-­Free­-Flow (AFF) System
  • Audible alarm to indicate errors or Pump Set loading conditions
  • Orientation­-independent design eliminates need for drip chamber on pump set
  • Sensor technology detects both upstream and downstream flow conditions
  • Continuous feed, intermittent feed, and flush capability
  • Auto­-prime feature reduces the need for time­-consuming manual priming
  • Hold feature with restart and Keep Tube Open (KTO) feature
  • Stores previous 72 hours of feeding and flushing history
  • 15 languages
  • Step-­by-­step prompts to guide programming and operation