Kangaroo 324

The Kangaroo 324 Enteral Feeding Pump provides accurate volumetric delivery. Retains flow rate, dose and volume delivered in memory for 24 hours after pump shut-off.

Description

The Sherwood Medical Kangaroo® Feeding Pump is a rotary peristaltic pump designed to regulate the flow rate of enteral feedings

Specifications
Type Infusion:  Device Volumetric 
Pumping Mechanism: Rotary Peristalic 
Pump Set Required: Sherwood Medical Kangaroo® Feeding Pump Administration Sets 
Operating Range: 

Rate: 1­-50 ml/hr in 1 ml/hr increments 50-­300 ml/hr in 5 ml/hr increments
Dose:5 2000 ml in 5 mi increments
Volume Delivered: 1­-9999 ml
Memory of Settings When Pump is Off 24 hours for all settings
Accuracy: flow ±10% of selected rate
Occlusion Pressure: 12 psi (82.7 kPa) Nominal
Alarms:

  • Flow Error
  • Low Battery
  • Hold Error
  • No Set
  • System Error
  • Dose Delivered
Battery: 

• Rechargeable
• Automatically charges when pump is plugged into wall outlet. Recharge time for fully discharged               battery is approximately 15 hours.
• A new battery, when fully charged, will operate for approximately 24 hours at 125 milliliters per hour.

Dimensions Approximately: 7.25" High x 5.5" Wide x 4.25" Deep
Weight: Approximately 3.625 Ibs.
Case Material:  ABS Plastic ­FIRE RESISTANT
Power Requirements: 120V, 60 Hz, 1 Amp
   
   
Features
  • Portable operation-­built in battery
  • Wide flow rate range
    1­-5 ml/hr in 1ml/hr increments
    50-­300 ml/hr in 5 ml/hr increments
  • Small and lightweight for easy transport
  • Touch panel operation
  • Shut off and activation of audible and visible alarms when:
    • feeding container is empty
    • feeding tube becomes occluded
    • drop sensors blocked
    • battery low (alarms 15 minutes before pump shut off)
    • unit is left on "HOLD" longer than 2 1/2 minutes
    • pump set is improperly loaded
    • self diagnosed problem which requires technical servicing
    • pump motor malfunction
  • Automatic systems check after pump has been turned ON
  • Large LED display
  • Removable power cord
  • Low occlusion pressure. Pump will not overcome back pressure greater than approximately 12psi (82.7kPa)
  • Integral pole clamp
  • Use of safety interlock pump sets to prevent free flow from inadvertent misload of pump set onto pump
  • Pump running indicator lights
  • DC motor with intermittent operation
  • Adjustable alarm volume
  • Memory of flow rate, dose, and volume delivered settings for 24 hours after pump shut off