CADD­-Solis

The CADD-Solis ® ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD­-Solis ® pump is indicated for intravenous, intra­-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion, patient­controlled demand doses, or both (such as patient-­controlled analgesia).

Description

The CADD-Solis ® ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD­-Solis ® pump is indicated for intravenous, intra­-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion, patient­controlled demand doses, or both (such as patient-­controlled analgesia).

Specifications
Dimensions:​  1.6" W X 4" H X 5" L 
Weight:​  1.31 lbs. (21 oz), including 4 AA batteries 
Resolution:​ Medication Cassette Reservoir or CADD ™ Administration Set, 0.050 ml per pump stroke nominal 
Moisture Protection:​ Splash-­proof (IPX4) per IEC 60529
Power:​ 4 AA alkaline batteries; AC adapter; rechargeable battery pack
Internal Memory Backup:​ The internal memory backup battery utilizes lithium manganese dioxide technology. It charges whenever the pump is powered on and has a 10 month memory capacity once it has been charged for 250 hours at 68°F
System Operating Temperature:​ 36°F to 104°F (2°C to 40°C)
System Storage Temperature: -​­4°F to 140°F (-­20°C to 60°C)
Atmospheric Pressure:​ 70 kPa (10.2 psi) to 106 kPa (15.4 psi)
Maximum Infusion Pressure: ​ 27.0 psi
Maximum Time to Occlusion Alarm: ​ 0.1 ml/hr – 122 min; 10 ml/hr – 122 sec
Bolus Volume at Occlusion​Alarm Pressure: 0.107 ml reservoir at 0.01 ml/hr: o 0.116 ml reservoir at 10 ml/hr:
Accuracy:​ ± 6% (nominal)
High Pressure Alarm Threshold: ​18 ± 9 psi
Air Detector Alarm: Sensitivity:
Low – Single bubble greater than 400 uL
High – Single bubble greater than 150 uL
Accumulated Air – Greater than 1 ml air over 15 minutes (nominal)
Maximum Volume Infused Under Single­Fault Conditions: ​CADD ® administration set: 0.15 ml
Delivery Rate during priming:​ Approx. 250 ml/hr
Alarm disabled during priming:​ Air-­in-­Line
Delivery Specifications:  
Reservoir Volume:​ 0 to 9999; programmable in 1 ml increments, displayed in 0.1 ml increments
Units:  ​milliliters (ml), milligrams (mg), micrograms (mcg)
Concentration: Mg/ml: 0.1, 0.2, 0.3, 0.4, 0.5, 1, 2, 3, …15, 20, 25, …100
Mcg/ml: 1, 2, 3, …15, 20, 25, …100, 200, 300, 400, 500
(Default = 1)
Continuous Rate:​ 0-­30 ml/hr (or the mg or mcg equivalent)
PCA Dose: 0 to 20 ml (or the mg or mcg equivalent)
Delivery rate (Continuous Rate + PCA Dose): programmable from 40­-175 ml/hr
PCA Dose Lockout:​ 1 minute to 24 hours in the following increments:
  1 minute for values between 1 and 20 minutes
5 minutes for values between 20 minutes and 24 hours
Max Doses per Hour:  ​1 – 60 doses 
Delivery Limit Amount:​ 0.1 to 1000 ml (or mg or mcg equivalent) in increments of
  0.01 ml from 0.01-­0.5 ml
0.5 ml from 0.5­-100 ml
1.0 ml from 100-­1000 ml
Given: ​0 to 99,999.99 in 0.01 unit increments
Clinician Bolus: 0 ­- 20 ml (or mg or mcg equivalent)
Delivery rate (Continuous Rate + Clinician Bolus): programmable from 40­-175 ml/hr
Alarms/Alerts:
Battery depleted while delivering Reservoir volume empty
Battery removed while delivering Pressure sensor faulty
Battery unusable while delivering Downstream occlusion
Rechargeable battery end of life Upstream occlusion
Disposable detached while delivering Air in Line detected
Disposable attached improperly Remote dose cord key stuck
Disposable type high flow administration set Key stuck
Disposable type invalid Stop mode reminder
Disposable locked but not latched Pump was automatically stopped
   
Features
  • Small and lightweight pump for convenience
  • Programmed intermittent bolus
  • Innovative therapies and task-­based menus
  • Scalable platform
  • User-­friendly interface
  • Graphical reports
  • Safety software