CADD Prizm VIP 6000 Series

The CADD-­Prizm® ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD­-Prizm® pump is indicated for intravenous, intra-­arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion. PCA delivery is intended for therapies that require a continuous rate of infusion, patient­controlled demand doses, or both (such as patient-­controlled analgesia). Continuous delivery allows the infusion of drug at a constant, programmed rate while intermittent delivery allows the infusion of a specific volume of drug at a regular, programmed interval. The final delivery program, TPN (Total Parenteral Nutrition), allows the infusion of nutritional solutions or other fluids, with optional tapering at the beginning and end of infusion.

Description

The CADD-­Prizm® ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD­-Prizm® pump is indicated for intravenous, intra-­arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion. PCA delivery is intended for therapies that require a continuous rate of infusion, patient­controlled demand doses, or both (such as patient-­controlled analgesia). Continuous delivery allows the infusion of drug at a constant, programmed rate while intermittent delivery allows the infusion of a specific volume of drug at a regular, programmed interval. The final delivery program, TPN (Total Parenteral Nutrition), allows the infusion of nutritional solutions or other fluids, with optional tapering at the beginning and end of infusion.

Specifications
Dimensions:​  1.7" W X 4.1" H X 5.6" L 
Weight:​ 1.25 lbs. (20 oz), including 9 volt battery and empty medication reservoir 
Resolution:​  Medication Cassette Reservoir or CADD ™ Administration Set, 0.050 ml per pump stroke nominal 
Power:​ 9 volt alkaline or lithium battery such as DURA­CELL ® Alkaline MN 1604 or ULTRALIFE® Lithium U9VL; CADD ™ External Power Source (EPS) Power Pack (reorder number 21-­3801); AC Adapter.
Battery Capacity:​ The expected life of a 9 volt battery is 12 hours at 100 ml/hour, or approximately 5 days at 10 ml/day (nominal). It is recommended that a new 9 volt battery be kept available for replacement if necessary.
System Operating Temperature:​ +36°F to 104°F (2°C to 40°C)
System Storage Temperature: ​­-4°F to 140°F (­-20°C to 60°C)
Power Pack Charging Temperature:​ +50°F to 95°F (10°C to 35°C)
Accuracy:​ ± 6% (nominal)
Bolus Volume at Occlusion​Alarm Pressure: ​0.050 ml resolution administration
Sets/Medication Cassette Reservoirs:
High Pressure Alarm: ​18 ± 9 psi
Air Detector Alarm:  ​Single bubble greater than 0.100 ml (can be turned ON or OFF)
PCA Delivery Mode Specifications:  
Reservoir Volume: ​1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments (Default: 1 ml)
Units: ​ milliliters (ml), milligrams (mg), micrograms (mcg); (Default: milligrams)
Concentration: Mg/ml: 0.1, 0.2, 0.3, 0.4, 0.5, 1, 2, 3, 4, 5, 10, 15, …95, 100
Mcg/ml: 1, 2, 3, 4, 5, 10, 15, …95, 100, 200, 300, 400, 500
(Default = 1)
Continuous Rate:​ 0­-30 ml/hr (or the mg or mcg equilavent); Default: 0 mg/hr
Demand Dose: 0 to 9.9 ml (Default: 0 mg/hr)
Delivery rate (Continuous Rate + Demand Dose): 125 ml/hr nominal
Demand Dose Lockout:​ 5 minutes to 24 hours in the following increments:
  1 minute for values between 5 and 20 minutes
5 minutes for values between 20 minutes and 24 hours
(Default: 5 minutes)
Max Doses per Hour: ​1 – 12 doses in 1 dose increments (will also be limited by the
Demand Dose Lockout value); Default: 1
Demand Doses Given:​ 0 to 999
Demand Dose Attempts:​ 0 to 999
Given:​ 0 to 99999.99 in 0.01 unit increments
Clinician Bolus: 0.1 ml to 20.00 ml (or mg or mcg equivalent)
Delivery rate (Continuous Rate + Clinician Bolus): 125 ml/hr nominal
Continuous Delivery Specifications:  
Reservoir Volume: ​ 1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments Default: 1 ml
Continuous Rate​: 0.1 – 350 ml/hr in the following increments:
  0.1 for values between 0.1 and 100
1 for values between 100 and 350
Default: 0 ml/hr
*Use 9 volt battery for rates up to 250 ml/hr; use power pack or AC adapter for rates up to 350 ml/hr.
Given:  ​0 to 99999.9 in 0.1 unit increments
TPN Delivery Mode Specifications:  
Reservoir Volume:  ​10.0 to 9990 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments (Default: 10 ml)
Infusion Volume:  ​10 – 9990 ml in 10 ml increments (Default: 10 ml)
Infusion Period:  ​0 hrs 10 min - 99 hrs 50 min in 10 minute increments (Default: 1 hr)
Taper-­Up Period:  ​0 hrs 0 min - 99 hrs 40 min in 10 minute increments (Default: 0 hr)
Taper-­Down Period: ​0 hrs 0 min - 99 hrs 40 min in 10 minute increments (Default: 0 hrs 0 mins)
Plateau Rate: ​ Calculated by pump; 10 - 350 ml/hr
KVO Rate:  ​Calculated by pump: 1/10 of Plateau Rate up to 5 ml/hr
Given: ​ 0 to 99999 in 0.1 unit increments
Intermittent Delivery Specifications:  
Reservoir Volume:​ 1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments (Default: 1 ml)
Dose Volume: ​0.1 – 1000 ml in the following increments:
  0.1 for values between 0.0 and 100
1 for values between 100 and 1000
Default: 0.0 ml
Dose Duration:​ 1 min ­- 24 hrs in the following increments:
  1 minute for values between 1 min and 10 min
5 minutes for values above 10 min
Default: 30 min
*Duration is limited by Dose Volume so that rate does not exceed 350 ml/hr. Use 9 volt battery for rates up to 250 ml/hr; use power pack or AC adapter for rates up to 350 ml/hr.
Dose Cycle:  ​10 min ­- 96 hrs in 5 minute increments (Default: 4 hrs)
KVO Rate: ​0 -­ 10 ml/hr in 0.1 ml/hr increments (Default: 0 ml/hr)
Next Dose Start Time: ​10 min -­ 96 hrs or Immediate; programmable in 10 minute increments (Default: Immediate)
Given: ​0 to 99999.9 in 0.1 unit increments
Time Remaining: ​Dose and Cycle display in 1 minute increments
Options:  
Immediate Taper-Down (TPN):  ​0 to time remaining in the infusion period, in 10 minute increments
AutoLock: ​Not In Use, LL1, or LL2
Time: ​00:00 to 23:59
Event Log: ​0 – 500 events
Extended History: ​Up to 48 hours in 1 hour increments
Alarms/Alerts:
Low battery power High delivery pressure
Depleted battery power Air in line
External power source low Air detector faulty or detached
Faulty Air detector port cover detached
Depleted Delivery too slow
Pump stopped Key stuck
Pump fault Cassette detached or unlocked
Low reservoir volume Print failure
Features
  • Easy to read display
  • Reliable and Durable
  • Programming flexibility
  • Air-­in­-Line detector
  • Event memory log
  • Lock levels for patient safety