CADD Legacy Plus

The CADD-­Legacy™ PLUS ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD-­Legacy™ PLUS pump is indicated for intravenous, intra­-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous or intermittent rate of infusion

Description

The CADD-­Legacy™ PLUS ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD-­Legacy™ PLUS pump is indicated for intravenous, intra­-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous or intermittent rate of infusion

Specifications
Dimensions:​  1.6" W X 3.8" H X 4.4" L 
Weight:​  0.86 lbs. (13.8 oz), including 2 AA batteries and empty medication reservoir 
Resolution: ​Medication Cassette Reservoir or CADD ™ Administration Set, 0.050 ml per pump stroke nominal
Moisture Protection:​ Splash­-proof (IPX4)
Power:​ Two AA alkaline batteries; AC Adapter
Battery Capacity:​ The expected life of 2 AA batteries is 15 hours at 125 ml/hour, or approximately 14 days at 10 ml/day (nominal). It is recommended that 2 new AA batteries be kept available for replacement if necessary.
Accuracy: ​± 6% (nominal). At low infusion rates, this accuracy may not be achieved for short periods. During the total infusion time, the accuracy averages out.
Maximum Infusion Pressure:​ 40.0 psi
Maximum Time to Occlusion Alarm:​ CADD® Administration Set with anti­siphon valve: 2.0 hours
Bolus Volume at Occlusion Alarm Pressure: ​0.050 ml resolution CADD ® Administration Sets/Medication Cassette™ Reservoirs:
Operating Temperature: ​+35°F to 104°F (2°C to 40°C)
Storage/Transportation Temperature:​­ -4°F to 140°F (­-20°C to 60°C)
High Pressure Alarm:​ 26 ± 14 psi
Air Detector Alarm: Single bubble Low Sensitivity: > 0.250 ml
Single bubble High Sensitivity: > 0.100 ml
Multi­bubble: 1.0 ml nominal
Maximum Volume Infused under Single Fault Condition: ​CADD® Administration Set with anti­-siphon valve: 0.2 ml
Delivery Rate during priming:  ​Approximately 180 ml/hr 
Alarm Disabled during priming:​  Air Detector 
Continuous Delivery Specifications:  
Reservoir Volume:​ 1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments (Default = 1.0 ml)
Continuous Rate: ​0.1 to 125 ml/hr; programmable in 0.1 ml/hr increments (Default = 0.0 ml)
Given:​ 0 to 99999.95 in 0.05 ml increments
Intermittent Delivery Specifications:  
Reservoir Volume:​ 1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments (Default = 1.0 ml)
Dose Volume: ​0.1 to 1000 ml programmable in 0.1 ml increments (Default = 0.0 ml)
Dose Duration: ​ Default = 30 min. Duration lasts 1 min to 24 hrs in the following increments:
  1 minute for values between 1 min and 10 min
5 minutes for values above 10 min
*Duration is limited by Dose Volume so that rate does not exceed 125 ml/hr
Dose Cycle: ​10 min to 96 hrs in 5 minute increments. Default = 4 hrs
KVO Rate: ​0 to 125 ml/hr in 0.1 ml/hr increments. Default = 0 ml/hr
Given:​ 0 to 99999.95 in 0.05 ml increments
Dose Remaining: ​1 minute increments
Alarms/Alerts:
Low battery power Disposable not attached when run attempted
Depleted battery power Motor locked
Battery dislodged Upstream occlusion
Pump stopped Reservoir volume empty
Pump fault Program incomplete
Low reservoir volume Key stuck
High delivery pressure Disposable detached
Air in line  
   
Features
  • Easy to read display
  • Cassette detection
  • Occlusion detection
  • Reliable and Durable
  • Programming flexibility
  • Air­-in-­Line detector with Off, High­ and Low­sensitivity settings
  • ON/OFF key allows pump to be put in a low-power state to save battery life
  • Event memory of the last 1,000 events in the pump's history