CADD Legacy 1

The CADD­-Legacy™ 1 ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD-­Legacy™ 1 pump is indicated for intravenous, intra­-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion.

Description

The CADD­-Legacy™ 1 ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD-­Legacy™ 1 pump is indicated for intravenous, intra­-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion.

Specifications
Dimensions:​  1.6" W X 3.8" H X 4.4" L 
Weight:​ 0.86 lbs. (13.8 oz), including 2 AA batteries and empty medication reservoir
Resolution:​ Medication Cassette Reservoir or CADD ™ Administration Set, 0.050 ml per pump stroke nominal
Moisture Protection:​ Splash­-proof (IPX4)
Power:​ Two AA alkaline batteries; AC Adapter
Battery Capacity:​ The expected life of 2 AA batteries is 15 hours at 125 ml/hour, or approximately 14 days at 10 ml/day (nominal). It is recommended that 2 new AA batteries be kept available for replacement if necessary.
Accuracy: ​± 6% (nominal). At low infusion rates, this accuracy may not be achieved for short periods. During the total infusion time, the accuracy averages out.
Maximum Infusion Pressure:​ 40.0 psi
Maximum Time to Occlusion Alarm:​ CADD® Administration Set with anti­siphon valve: 2.0 hours 
Bolus Volume at Occlusion Alarm Pressure:​  0.050 ml resolution CADD ® Administration Sets/Medication Cassette™ Reservoirs:
Operating Temperature: ​+35°F to 104°F (2°C to 40°C)
Storage/Transportation Temperature: ​­-4°F to 140°F (­-20°C to 60°C)
High Pressure Alarm:​ 26 ± 14 psi
Air Detector Alarm: Single bubble Low Sensitivity:​> 0.250 ml
Single bubble High Sensitivity:​> 0.100 ml
Multi­bubble:​1.0 ml nominal
Maximum Volume Infused under Single Fault Condition: ​ CADD® Administration Set with anti­siphon valve: 0.2 ml
Delivery Rate during priming:​ Approximately 180 ml/hr
Alarm Disabled during priming:​ Air Detector
Continuous Delivery Specifications: Reservoir Volume: 1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments (Default = 1.0 ml)
Continuous Rate: 1 to 3000 ml/24 hr; programmable in 1 ml/24 hr increments (Default = 1.0 ml)
0 to 99999.95 in 0.05 ml increments
Alarms/Alerts:
  • Low battery power
  • Depleted battery power
  • Battery dislodged
  • Pump stopped
  • Pump fault
  • Low reservoir volume
  • High delivery pressure
  • Air in line
  • Disposable not attached when run attempted
  • Motor locked
  • Upstream occlusion
  • Reservoir volume empty
  • Key stuck
  • Disposable detached
   
   
   
Features
  • Easy to read display
  • Cassette detection
  • Occlusion detection
  • Air-­in-­Line detector with Off, High­-and-Low­sensitivity settings
  • ON/OFF key
  • Event memory
  • Reliability
  • Programming flexibility